SLT_424 CalVer Assessment

Use this template for Calibration Verification when replicates are performed. The Mean of the replicates is compared with the allowable limits, and Target for acceptability and % Recovery.


I like this template when using previously tested samples (usually frozen), and re-testing for the required 6-month Calibration Verification study.


Calibration Verification:

  • Enter Target Value and Allowable Error Limit %, and Target Limits are automatically calculated
  • Enter Observed Values, a Mean is calculated and compared to the Target Limits for % Recovery
  • Interpretation of "Within Range" or "Exceeds Range" is given
  • Verify Analyte Measurement Range and Document Replicate Reproducibility